Covidien Force FX-CS Electr...
Covidien Force FX-CS Electrosurgical Generator W/ E6008 E6009 FootControl~36527
Item specifics
- Used
- Model
- FX-CSMPN
- FX-CS
Product details
Manufacturer | Covidien |
Item: | 2014 Covidien Force FX-CS Electrosurgical Generator W/ E6008 Monopolar & E6009 Bipolar Foot Controls |
Model/Cat # | FX-CS |
Serial No. | S4G1320AX |
Made In: | USA |
Date of Manufacture: | 2014-07 |
Voltage: 100-240V 50-60hz 300W
Short description...
- 1. Sale is for imaged items only.
- 2. Display signs of some usage.
- 3. Has been tested at our end & is in good functionality.
A. Covidien Force FX-CS ESU:
*Unit has some scratches & scuffs
*No dents or cracks have been observed
*At the left side of the panel 2 tape pieces have been applied by the facility
*Softpads: All intact & functional
*Handle: Intact
*Top of the console: Seems to have some scratches due to prior usage
*Bottom of the console: Has some scratches but all feet are intact
*Rear panel enjoys ports for Monopolar (1&2) Bipolar & VGA/RS232
B. Valleylab Monopolar Foot Control Ref E6008 |Lot# 701818 | DOM: 2007-05
*Cable: 15-ft long | Has some scratches but no cuts, tears or leakages observed
*Strain reliefs: Tear-free
*Connector: 4-pin connector; No bent or broken pins observed
*Foot pedal: Occasional scratches & chips of paint loss may be there at different places
*Response: Both pedals are responsive
C. Valleylab Bipolar Foot Control Ref E6009 | Lot# 692385 | DOM: 2007-03
*Cable: 15-ft long | Has some scratches but no cuts, tears or leakages observed
*Strain reliefs: Tear-free
*Connector: 3-pin connector; No bent or broken pins observed
*Foot pedal: Looks good
*Response: Responding to commands
*Outputs:
-Monopol: 300W / 300 Ohm
-Bipol: 70W / 100 Ohm
-Frequency: 390 KHz
-Interval: 10s / 30s
C. Testing
*Unit turns on > Passed selftest > Responding to commands > Responding to Hand control/Foot control
*Note: Not labelled as patient ready | Advised to have PM & calibration before clinical use
*No manual, consumables or disposables with device
*Cosmetically in good shape
*Please review images & check for compatibility before bidding for the item
Package includes:-
Covidien Force FX-CS Electrosurgical Generator
Valleylab Monopolar Foot Control Ref E6008 |Lot# 701818 | DOM: 2007-05
Valleylab Bipolar Foot Control Ref E6009 | Lot# 692385 | DOM: 207-03
Power Cord
*Items shown in photos are the items that will ship. If it is not shown in the photos, it is not included*
- 9. Shipping Weight : 35 lbs
Stock ID : 36527 {P.O 1200} (WH22)
Comments:
Kindly enlarge the images and see them with zooming.
Concern or query: Ask a Question
Condition: USED, please refer to images for cosmetic condition.
- 1. Overall Condition: As seen in images : Good
- 2. Physical Condition: Please review images before bidding : Good
- 3. Working Condition: Has been tested at our end & is in good functionality....Will receive item as pictured ……………(offered AS IS)
- 4. “As a whole; the item has not been tested technically or professionally at our end- unable to test --- being sold STRICTLY AS IS! Biomed check is strongly recommended before clinical use!”
- 5. Completeness: (with everything seen in the pictures. Please look at all pictures and go through detail description before bidding.)